ELECTRONIC BLOOD PRESSURE MONITOR Instruction Manual MODEL: ABP802
The ABP802 Blood Pressure Monitor uses the oscillometric method of blood pressure measurement. Measurement Automatic Electronic Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure, and pulse rate on adults each time, with an arm cuff around the left upper arm. The expected life of the product is 5 years. The product complies with the electromagnetic compatibility requirement of EN60601-1-2 and safety standards of EN60601-1 and performance of EN1060-1. EN1060-3 as specified in EEC directive 93/42/EEC.
Patient must follow doctor's instruction and should not perform self-judgment and self-treatment by the measuring result. Self-diagnosis of measured results and treatment are dangerous. The device should not be used to judge illness, first aid, and continuously measuring. This device cannot be used for patient transport and surgical care. It can be used in a household or fixed place only. Please press "on/off" button to stop the device when you feel uncomfortable, or if the air is inflating abnormally without stopping.
Do not let a child below 12 years old and people who cannot express their intention use this. When it is used by the people of 12~18 years old, they should be accompanied by an adult.
Do not use the unit for a purpose other than measuring blood pressure.
Do not use a mobile phone around the device. Please do not use the device around a magnetic field.
Do not disassemble, repair, or remodel the main unit or the wrist cuff of the blood pressure monitor. This will cause the unit to malfunction.
Requests from Manufacturer
Make sure the connection tubing is not kinked before measuring to avoid any injury to patient. |
For any patient, do not measure more than 3 times continuously. Wait for at least 5 minutes between any two measurements; otherwise it will cause extravasated blood. |
Do not measure your blood pressure over 6 times each day. |
Do not apply the cuff over a wound as this can cause further injury. |
Do not measure on the arm on the same side as a mastectomy, otherwise it could cause injury. |
Observe the air pressure value on the LCD display. |
When measuring, it should not exceed 299 mmHg, otherwise please press the "on/off" button to stop. |
Do not use force to bend the arm cuff or the air tube. |
Do not knock or drop the main unit. |
Always use the accessories specified in the manual, the use of other parts not approved by the manufacturer may cause faults or injuries. |
For service information, parts list, etc., please contact the dealer. |
ABOUT BLOOD PRESSURE
1. What is blood pressure? Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts. Diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg). One's natural blood pressure is represented by the fundamental pressure, which is measured first thing in the morning while one is still at rest and before eating.
2. What is hypertension and how is it controlled? Hypertension, an abnormally high arterial blood pressure, if left unattended, can cause many health problems including stroke and heart attack. Hypertension can be controlled by altering lifestyle, avoiding stress, and with medication under a doctor's supervision. To prevent hypertension or keep it under control: * Do not smoke * Exercise regularly * Reduce salt and fat intake * Have regular physical checkups * Maintain proper weight
3. Why measure blood pressure at home? Blood pressure measured at a clinic or doctor's office may cause apprehension and produce an elevated reading, 25 to 30 mmHg higher than that measured at home. Home measurement reduces the effects of outside influences on blood pressure readings, supplements the doctor's readings, and provides a more accurate, complete blood pressure history.
4. W.H.O. blood pressure classification Standards for assessment of high blood pressure, without regard to age, have been established by the World Health Organization (WHO), and are shown in chart to the right.
5. Blood pressure variations. An individual's blood pressure varies greatly on a daily and seasonal basis. It may vary by 30 to 50 mmHg due to various conditions during the day. In hypertensive individuals, variations are even more pronounced. Normally, the blood pressure rises while at work or play and falls to its lowest levels during sleep. So, do not be overly concerned by the results of one measurement. Take measurements at the same time every day using the procedure described in this manual to determine your normal blood pressure. Many readings give a more comprehensive blood pressure history. Be sure to note date and time when recording your blood pressure. Consult your doctor to interpret your blood pressure data.
PRECAUTIONS BEFORE USE
1. If you are taking medication, consult with your doctor to determine the most appropriate time to measure your blood pressure. NEVER change a prescribed medication without first consulting with your doctor.
2. For people with irregular or unstable peripheral circulation problems due to diabetes, liver disease, hardening of the arteries, etc., there may be fluctuation in blood pressure values measured at the upper arm versus at the wrist.
3. Measurements may be impaired if this device is used near televisions, microwave ovens, X-ray, mobile phone equipment, or other devices with strong electrical fields. To prevent such interference, use the monitor at a sufficient distance away from such devices or turn them off.
4. Before using, you should wash your hands.
5. Do not measure on the arm which is being used by monitoring ME Equipment. Otherwise it could cause loss of function.
6. Consult your doctor if unexpected readings are obtained. Also please refer to "TROUBLESHOOTING" section of the manual.
7. The reading is probably a little lower than measured in the hospital due to the steady mood at home.
8. Cuff pressure range is 0-299mmHg
FEATURES OF THE PRODUCT
1. Memory can store 90 measurements.
2. Large and clear LCD display.
3. WHO blood pressure classification display.
4. Easy to use. Press a button to automatically measure and record the measurement values and measurement time.
5. Automatically turns off (within 1 minute) to save power.
INSERT OR REPLACE BATTERIES
1. Remove the battery cover.
2. Insert new batteries into the battery compartment as shown, taking care that the polarities (+) and (-) are correct.
3. Close the battery cover. Use only LR6, AA batteries.
Dispose of dead batteries to the authorized collecting party according to the regulation of each individual territory.
CAUTION
* Insert the batteries as shown in the battery compartment. If not, the device will not work.
* When (LOW BATTERY mark) blinks on the display, replace all batteries with new ones. Do not mix old and new batteries. It may shorten the battery life, or cause the device to malfunction. (LOW BATTERY mark) does not appear when the batteries run out.
* Battery life varies with the ambient temperature and may be shorten at low temperatures.
* The batteries may leak and cause a malfunction.
* Use the specified batteries only. The batteries provided with the device are for testing monitor performance and may have a shorter life.
* Used batteries may leak and damage the main unit. Pleases observe the following points.
- lf you are not going to use the unit for a long period of time (approximately three months or more), remove the batteries.
- Replace worn batteries with their polarities in the correct direction.
TIME AND VOICE ON/OFF OF SYSTEM SETUP
1. Press "SET" key to turn on.
2. Press and hold "SET" key until the year number displays and flashes on LCD to enter the settings mode.
3. Press the "MEM" key to adjust the year, then press "SET" key again to save your setting and enter the month setting mode.
4. Press "MEM" key to adjust the month. Following the same steps to adjust date/hour/minute/(Voice on/off) until setting completed ("ON" is the On, "OF" is Off) Non-talking model does not have this function.
UNIT CONVERSION mmHg/kPa DISPLAY
The monitor has two kinds of blood pressure display units mm Hg(mmHg), kPa (kPa). (mmHg is the factory default). While the monitor is off, press and hold the ON/OFF button for 5 seconds. The monitor will boot up and measure showing the other unit. (If it was in mmHG it will change to kPa and vice versa.) Also the stored values will be shown in the new unit.
WHO BLOOD PRESSURE CLASSIFICATION DISPLAY
Diastolic blood pressure reference material: Journal of Hypertension 1999. Vol 17 No.2
ATTACHING THE ARM CUFF
1. Wrap the arm cuff around the upper arm, about (2-3) cm above the elbow, as shown. Place the cuff directly on the skin, as clothing may cause a faint pulse, and result in a measurement error.
2. Constriction of the upper arm, caused by rolling up a shirtsleeve, may prevent accurate readings.
3. Secure the arm cuff with Velcro Strip in such a way that it lies comfortably and is not too tight. Lay the arm on the table (palm upwards) so that the arm cuff is at the same height as the heart. Make sure that the tube is not kinked.
4. Measure your arm circumference for cuff selection. Please refer to "SPECIFICATIONS" section.
HOW TO TAKE PROPER MEASUREMENTS
For the most accurate blood pressure measurement please adhere to the following guidelines:
1) Comfortably seated
2) Legs uncrossed
3) Feet flat on the floor.
4) Back and arm supported
5) Middle of the cuff at the same level as the heart
* Remain still and keep quiet during measurement. Relax as much as possible and remain from talking during the measurement process.
* Measure your blood pressure at about the same time every day. Do not measure right after physical exercise or a bath. Take a rest for twenty or thirty minutes before taking the measurement.
* The reading could be affected by the following: Within in an hour after dinner, after having wine, coffee, red tea, sports, bathing; talking, being nervous, being in unsteady mood, bending forward, moving, room temperature dramatically changing during measuring; in moving vehicles, long continuous measuring.
HOW TO MEASURE BLOOD PRESSURE
1. Setup the arm cuff to your upper arm as previous section of "ATTACHING THE ARM CUFF"
2. Press the "ON/OFF" button, all icons appear for two seconds on DISPLAY, then switch to measurement, and displays "0" or the last measurement recorded.
3. To start the measurement, the cuff in the strap will automatically inflate. The Heart Icon will flash on the LCD. When complete, the LCD will display measurement results.
READ MEMORY
Press the "MEM" button to view the last measurement recorded. Press the "MEM" button again to go through older records. Press the "SET" button to go to more recent records. The number next to "M" on the screen indicates which record it is. The larger the number is, the older the record.
DELETE MEMORY
Press and hold the "MEM" button for 5 seconds to delete all of the records in memory. The LCD will display "No" after the records have been deleted.
CARE AND MAINTENANCE
To keep your digital blood pressure monitor in the best condition and protect the unit from damage, follow the directions listed below:
-Keep the monitor in the storage case when not in use.
- Do not fold the arm cuff too tightly. The fabric fastener could touch the inner surface of the arm cuff and damage it.
-Clean the monitor and cuff with a soft dry cloth.
- Do not use any abrasive or volatile cleaners.
CAUTION
* Do not submerge the device or any of the components in water.
* Do not subject the monitor to extreme hot or cold temperatures, humidity, or direct sunlight.
* Store the device and the components in a clean, safe location.
* Do not subject the monitor to strong shocks, such as dropping the unit on the floor.
* Remove the batteries if the unit will not be used for three months or longer. Always replace all the batteries with new ones at the same time.
* This product is designed for use over an extended period of time; however, it is generally recommended that it be inspected and calibrated every two years to ensure proper function and performance.
SPECIFICATIONS
Measuring Method | Oscillometric Measurement |
Indication | Digital LCD display |
Measuring Range: | Pressure:(30~280)mmHg Pulse:(40~199)Beat/min |
Accuracy: | Static Pressure: ±3mmHg Pulse: ±5% |
Memory: | 90 Readings |
Power supply: | 4 x 1.5V Batteries (LR6 or AA) Using alkaline batteries will provide over 200 measurements. |
Operating condition: | +5°C ~ +40°C 15%RH~93%RH Atmospheric pressure: 70kPa~106kPa |
Storage condition: | -20°C ~ +55°C 0%RH~93%RH Atmospheric pressure: 50kPa~106kPa |
Dimensions: | Approx.: 140(W)X110(H)X48(D)mm |
Weight: | Approx.: 450g, excluding batteries |
Classification | Type BF |
Upper arm circumference | (22~32)cm |
AC adapter | INPUT: 100-240V~ 50/60Hz, 0.3A OUTPUT: 6V 500mA |
* Specifications may be changed without notice in the event of improvements being made.
1. Type of protection against electric shock: INTERNALLY POWERED EQUIPMENT.
2. Degree or protection against electric shock: TYPE BF APPLIED PART.
3. Mode of operation: CONTINUOUS OPERATION.
4. Equipment not suitable for category AP&APG equipment use in presence.
STATEMENT
The system might not meet its performance specifications if stored or used outside the temperature and humidity as mentioned below:
Operating conditions: +5°C ~+40°C 15%RH~93%RH 70kPa~106kPa
Storage conditions: -20°C ~+55°C 0%RH~93%RH
TROUBLESHOOTING
If you have trouble in using the unit please check the following points first.
ERROR DISPLAY | POSSIBLE CAUSE | HOW TO CORRECT |
Nothing is displayed when you push the POWER button or battery icon flashes. | -No battery installed -Batteries are dead -The batteries are installed incorrectly |
-Insert batteries -Replace with new batteries -Insert battery in the correct polarities |
Cannot normally increase pressure | Check your arm cuff for any air leakage | Replace arm cuff with new one |
Inflate pressure is too high | Re-measure or send back to dealer to re-calibrate pressure | |
"E3" | Hand or body shaking while measuring | Keeping still and measure again |
Battery icon flashes on | Batteries have low power | Replace batteries and measure again |
The systolic pressure value or diastolic pressure value too high | -The arm cuff was held lower than your heart -The arm cuff was not attached properly -You moved your body or spoke while measuring |
Keep a correct position, remain still, and measure again |
The systolic pressure value or diastolic pressure value too low | -The arm cuff was held higher than your heart -You moved your body or spoke while measuring |
Keep a correct position, remain still, and measure again |